實驗室離心機的一些基礎維修方式
2020-04-15(1201)次瀏覽
(1)震動劇烈、噪音大
檢查(cha)是(shi)否(fou)有不平衡的問題存在、固(gu)定機(ji)器的螺(luo)帽(mao)松動(dong),如有則(ze)固(gu)緊;檢查(cha)軸(zhou)承(cheng)是(shi)否(fou)損壞或(huo)彎曲,如有則(ze)更換(huan)軸(zhou)承(cheng);機(ji)器外罩變形或(huo)位置不正確(que)發生摩擦,如有則(ze)對其進行調整。
(2)通電后電機不轉動
首先對電源(yuan)線(xian)(xian)、插頭、插座進行檢(jian)查(cha),如(ru)(ru)有損(sun)(sun)壞(huai)則(ze)應更換;如(ru)(ru)無問(wen)題則(ze)檢(jian)查(cha)波段開(kai)關或(huo)變(bian)阻器(qi)是否損(sun)(sun)壞(huai)或(huo)連線(xian)(xian)開(kai)脫(tuo)(tuo)(tuo)(tuo),如(ru)(ru)損(sun)(sun)壞(huai)或(huo)開(kai)脫(tuo)(tuo)(tuo)(tuo)則(ze)更換損(sun)(sun)壞(huai)元件,重新焊接連線(xian)(xian);如(ru)(ru)無問(wen)題則(ze)檢(jian)查(cha)電機(ji)線(xian)(xian)圈(quan)是否有斷(duan)脫(tuo)(tuo)(tuo)(tuo)或(huo)斷(duan)路(內(nei)部),如(ru)(ru)是連線(xian)(xian)圈(quan)斷(duan)脫(tuo)(tuo)(tuo)(tuo)可以(yi)重焊.如(ru)(ru)是線(xian)(xian)圈(quan)內(nei)部斷(duan)路則(ze)只能重繞線(xian)(xian)圈(quan)。
(3)電動機轉速達不到額定轉速
首先(xian)檢查軸承(cheng)(cheng),如(ru)軸承(cheng)(cheng)損壞(huai)則(ze)更換軸承(cheng)(cheng),如(ru)軸承(cheng)(cheng)內部缺油或(huo)污物(wu)太多則(ze)應(ying)(ying)清洗軸承(cheng)(cheng)并加注潤滑(hua)脂(zhi)。檢查整(zheng)流子(zi)(zi)表面是(shi)(shi)否(fou)有異常或(huo)電刷與整(zheng)流子(zi)(zi)轟(hong)面的配(pei)(pei)合是(shi)(shi)否(fou)吻合,如(ru)整(zheng)流子(zi)(zi)表面有異常,如(ru)有一層氧化物(wu)則(ze)應(ying)(ying)用細砂紙對其(qi)進行打磨;如(ru)整(zheng)流子(zi)(zi)與電刷的配(pei)(pei)合不吻合則(ze)應(ying)(ying)調整(zheng)到接觸良好的狀態;如(ru)無以上問題則(ze)檢查轉子(zi)(zi)線(xian)圈中是(shi)(shi)否(fou)有短(duan)路的現(xian)象,如(ru)有則(ze)應(ying)(ying)重(zhong)繞線(xian) 78
圈;檢查調速電(dian)阻W 是否可調,若(ruo)沒(mei)有損壞,測量可控硅SCR、SCR2的(de)導通(tong)電(dian)壓(ya)Ug是否屬于正(zheng)常(chang),若(ruo)正(zheng)常(chang)測量SCR1、SCR2的(de)導通(tong)情況,正(zheng)常(chang)時(shi)為(wei)正(zheng)向(xiang)導通(tong),反(fan)向(xiang)截止(zhi),反(fan)之(zhi),更(geng)換可控硅。
(4)整流子與電刷之間火花大
檢查整(zheng)(zheng)(zheng)流子(zi)表面(mian)是(shi)(shi)否平整(zheng)(zheng)(zheng),如表面(mian)不平整(zheng)(zheng)(zheng)可(ke)以(yi)用(yong)砂紙打磨,若(ruo)是(shi)(shi)電(dian)刷與整(zheng)(zheng)(zheng)流子(zi)接觸(chu)不良,則用(yong)細(xi)砂紙將其部(bu)(bu)位打磨,重新(xin)調整(zheng)(zheng)(zheng)電(dian)刷與整(zheng)(zheng)(zheng)流子(zi)之間的配合;若(ruo)是(shi)(shi)電(dian)刷問題,則只能更(geng)換電(dian)刷;如果以(yi)上無(wu)問題,則要檢查磁場(chang)線(xian)(xian)圈(quan)(quan)(quan)是(shi)(shi)否有局(ju)部(bu)(bu)短路(lu)或(huo)線(xian)(xian)圈(quan)(quan)(quan)對地,還是(shi)(shi)轉(zhuan)子(zi)線(xian)(xian)圈(quan)(quan)(quan)有局(ju)部(bu)(bu)短路(lu)或(huo)斷(duan)路(lu)。磁場(chang)線(xian)(xian)圈(quan)(quan)(quan)和轉(zhuan)子(zi)線(xian)(xian)圈(quan)(quan)(quan)無(wu)論出(chu)現哪一個壞(huai)都要重繞(rao)磁場(chang)線(xian)(xian)圈(quan)(quan)(quan)或(huo)重繞(rao)轉(zhuan)子(zi)線(xian)(xian)圈(quan)(quan)(quan)。
(5)離心室內不降溫
離(li)(li)心(xin)(xin)機(ji)制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)是(shi)(shi)(shi)采用壓(ya)縮式單級蒸發循環制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng),制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)過程(cheng)中必須要有(you)(you)兩個系(xi)統(tong)組成。其一是(shi)(shi)(shi)控(kong)制(zhi)(zhi)(zhi)(zhi)(zhi)系(xi)統(tong),其二是(shi)(shi)(shi)制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)系(xi)統(tong)。可(ke)見,離(li)(li)心(xin)(xin)室(shi)不降溫(即不制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng))有(you)(you)可(ke)能(neng)是(shi)(shi)(shi)控(kong)制(zhi)(zhi)(zhi)(zhi)(zhi)系(xi)統(tong)和制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)系(xi)統(tong)出現故(gu)障,先檢查(cha)(cha)確認是(shi)(shi)(shi)控(kong)制(zhi)(zhi)(zhi)(zhi)(zhi)系(xi)統(tong)還是(shi)(shi)(shi)制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)系(xi)統(tong)出現故(gu)障。如(ru)果(guo)是(shi)(shi)(shi)制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)壓(ya)縮機(ji)運轉而(er)離(li)(li)心(xin)(xin)室(shi)內不降溫,可(ke)能(neng)就是(shi)(shi)(shi)制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)系(xi)統(tong)漏氟(fu)或堵塞,就要對制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)系(xi)統(tong)查(cha)(cha)漏重新加氟(fu);如(ru)果(guo)制(zhi)(zhi)(zhi)(zhi)(zhi)冷(leng)(leng)(leng)壓(ya)縮機(ji)不運轉,就是(shi)(shi)(shi)控(kong)制(zhi)(zhi)(zhi)(zhi)(zhi)系(xi)統(tong)故(gu)障,先查(cha)(cha)離(li)(li)心(xin)(xin)室(shi)的溫度設定值、顯示改變是(shi)(shi)(shi)否正常,如(ru)果(guo)正常,查(cha)(cha)離(li)(li)心(xin)(xin)室(shi)溫度傳感(gan)器及插板各插頭及接線,是(shi)(shi)(shi)否接觸良好,壓(ya)縮機(ji)啟(qi)動(dong)器(PTC)是(shi)(shi)(shi)否損壞,壓(ya)縮機(ji)線圈(quan)斷線,逐一排查(cha)(cha)并解決。
最(zui)新資訊
-
共赴新征程 共書新傳奇 —— 湘立儀器 2026 新春年會凝心聚力啟新篇
金歲收官(guan)啟新(xin)章,喜樂相逢赴前程。2026 年新(xin)春,湖南湘立科學儀...
-
湘立攜“PRP”醫療離心機亮相第92屆中國國際醫療器械博覽會
第(di)92屆中國國際醫療器械博覽(lan)會(CMEF)于9月(yue)26-29日(ri)在廣州啟幕。...
-
湘立儀器:三渡危機,轉型鑄就離心機領域領軍者
從疫情(qing)時(shi)期的瀕臨倒閉,到如今榮獲“湖南省博士創新站(zhan)”認證,湖...
-
血小板衍生制品在創面愈合中的研究進展及湘立 PRP 專用離心機的技術支撐
一、血小板(ban)及其衍生制品在(zai)創面愈合中的作用與研(yan)究進展(zhan)血小板(ban)是直...
400 082 6822
服務熱(re)線:400 082 6822
聯(lian)系電話:133 9751 9184
公司傳真(zhen):0731-84863976
公司郵箱:xiangli@apvsfd.cn
公司地(di)址(zhi):湖南省長沙高新(xin)開(kai)發區桐梓坡西路229號(hao)麓谷(gu)國際工業(ye)園A1棟





400 082 6822 